Radiological interventions play an increasingly relevant role in cardiology. Due to the inherent risks of ionizing radiation, proper\ncare must be taken with monitoring and optimizing the dose delivered in angiograms to pose as low risk as possible to staff and\npatients. Dose optimization is particularly pertinent in teaching hospitals, where longer procedure times are at times necessary to\naccommodate the teaching needs of junior staff, and thus impart a more significant radiation dose. This study aims to analyze the\neffects of different protocol settings in routine coronary angiograms, from the perspective of a large tertiary center implementing\na rapid dose reduction program. Routine coronary angiograms were chosen to compare baseline levels of radiation, and the dose\nimparted before and after dose optimization techniques was measured. Such methods included lowering dose per pulse,\nfluoroscopic pulse rates, and cine acquisition frame rates. The results showed up to 63% reduction in radiation dose without\nadverse impact on clinical or teaching outcomes. A 10 fps/low and 5 pps/low setting was found to achieve maximum dose\noptimization, with the caveat that settings require incremental changes to accommodate for patient complexities.